Invacare IRC5LXO2AW Oxygen Equipment User Manual


 
SECTION 7—TROUBLESHOOTING GUIDE
Platinum™ 32 Part No 1134866
YELLOW or RED
light illuminated.
Alarm:
Continuous
On red light only.
* Only applies to
SensO
2
models.
1. Low oxygen purity.*
2. Kinked or blocked
tubing, cannula or
humidifier.*
3. Flowmeter set at 0.5
L/min. or less.*
4. Unit overheating due
to blocked air intake.
5. Insufficient power at
outlet.
6. Internal repairs
required.
1. Clean or Replace filters.
2. Inspect for kinks or
blockages. Correct, clean or
replace item. Once
corrected, turn power OFF
for 60 seconds and then
turn power back ON.
3. Check flowmeter is set to
1.0 L/min. or more. (Refer
to Recommended
Guidelines for Optimum
Performance on page 20.
4.a. Remove and clean
cabinet filters.
4b. Move concentrator at
least 7.5 cm (3 in) from
walls, draperies, and
furniture.
5. DO NOT use extension
cords. Move to another
electrical outlet or circuit.
6. Call Service Provider.
GREEN light
YELLOW Light
Flashing.
1. Internal repairs
required.
1. Call Service Provider.
Alarm:
RAPID
Beep....Beep...
Beep...Beep
Low Flow Alarm
1a. Kinked or blocked
tubing, cannula or
humidifier.
1b. Flowmeter set at 0.5
L/min. or less.
1a. Inspect for kinks or
blockages. Correct, clean or
replace item. Once
corrected, turn power OFF
for 60 seconds and then
turn power back ON.
1b. Reset flowmeter to
prescribed flowrate.
Flowrates less than1L/min.
are not recommended.
SYMPTOM: PROBABLE CAUSE: SOLUTION: