Invacare IRC5LXO2AW Oxygen Equipment User Manual


 
SECTION 6—MAINTENANCE
Platinum™ 28 Part No 1134866
SECTION 6—MAINTENANCE
NOTE:TheInvacareconcentratorsarespecificallydesignedto
minimizeroutinepreventivemaintenance.Inplaceswithhighdust
orsootlevels,maintenancemayneedtobeperformedmoreoften.
Onlyauthorizedorfactorytrainedpersonnelshouldperform
preventivemaintenanceontheconcentrator.Powershouldbe
disconnectedbeforebeginningpreventivemaintenanceonthe
Invacareconcentrator.
NOTE:Ataminimum,preventivemaintenanceMUSTbe
performedaccordingtothemaintenancerecordguidelines.Referto
PreventiveMaintenanceRecordonpage 30.
Routine Maintenance
NOTE:Thefollowingroutinemaintenanceshouldbeperformed
morefrequently.Seerecommendedintervalsineachsectionand
preventivemaintenancerecord.
Cleaning the Cabinet Filter
NOTE:Forthisprocedure,refertoFIGURE6.1onpage 29.
CAUTION
DO NOT operate the concentrator without the filters
installed.
NOTE:Therearetwocabinetfilters.Onefilterlocatedoneachside
ofthecabinet.
1. Removeeachfilterandcleanatleastonceaweek
dependingonenvironmentalconditions.
NOTE:Environmentalconditionsthatmayrequiremorefrequent
cleaningofthefiltersincludebutarenotlimitedto:highdust,
smoking,airpollutants,etc.
2. Cleanthecabinetfilterswithavacuumcleanerorwashin
warmsoapywaterandrinsethoroughly.