Platinum™ 22 Part No 1134866
3. Removethefilteraccessdoorlocatedonthesideofthe
concentrator.Insertaflatheadscrewdriverintheplate
grooveonthetopedgeofthefilteraccessdoorandgently
prythefilteraccessdooropen.RefertoDetail“Ain
FIGURE5.3.
4. Thehumidifieradapterisnexttotheinletfilter.Pullup
andremovethehumidifierbottleadapter.RefertoDetail
“B”inFIGURE5.3.
5. Attachittothehumidifierbyturningthewingnutonthe
humidifierbottlecounterclockwiseuntilitissecurely
fastened.RefertoDetail“C”inFIGURE5.3.
FIGURE 5.3 Attaching the Humidifier Bottle Adapter
DETAIL “A”
Plate Groove (Insert Flat
Head Screwdriver Here)
Filter Access Door
DETAIL “B”
Humidifier Bottle
Adapter
Humidifier Bottle
Adapter
Filter
Access
Door