Invacare IRC5LXO2AW Oxygen Equipment User Manual


 
SECTION 5—OPERATING INSTRUCTIONS
Platinum™ 24 Part No 1134866
FIGURE 5.5 Power Switch
Flowrate
NOTE:Forthisprocedure,refertoFIGURE5.6onpage 25.
1. Turntheflowrateknobtothesettingprescribedbyyour
physicianortherapist.
ƽ WARNING
DO NOT change the L/min. setting on the flowmeter
unless a change has been prescribed by your physician or
therapist.
NOTE:Toproperlyreadtheflowmeter,locatetheprescribed
flowratelineontheflowmeter.Next,turntheflowknobuntilthe
ballrisestotheline.Now,centertheballontheL/min.line
prescribed.
2. Iftheflowrateontheflowmeterever fallsbelow0.5L/min.
formorethanaboutoneminute,theLOWFLOWalarm
willbetriggered.Thisisarapidbeepingoftheaudible
alarm.Checkyourtubingoraccessoriesforblockedor
kinkedtubingoradefectivehumidifierbottle.Afterrated
flowisrestored,morethan0.5L/min.,theLOWFLOW
audiblealarmwillgooff.
On/Off (I/O)
Power Switch