Invacare HF2E540M9 Oxygen Equipment User Manual


 
SECTION 1—GENERAL GUIDELINES
Part No 1163156 7 HomeFill
®
Cylinder
ƽ WARNING
NEVERoilorlubricatethecylinderfittings.NEVERuseany
penetratingoilorlubricantsuchasWD40or3inOne.If
connectiondifficultyisexperiencedverifythatthecylinder
fittingisbeinginsertedstraightdownintothecompressor
coupler.Ifcontinueddifficultyisexperiencedcontactyour
supplierforassistance.
DONOTallowcylinderstotiporfall.Securegascylindersso
theycannotfall.Ensurecylinderisalwaysinuprightposition.
DONOTstorecylindersnearsourcesofheatorflame.
DONOTusewhilesleepingwithoutconsultingyour
healthcareprovider.
DONOTplacetheunitunderclothingwhileinuse.When
thisproductisinuseasmallamountofoxygenisvented.
Wearingtheunitunderclothingmaysaturatefabricswith
oxygenandcausethemtoburnrapidlyifexposedtosparks
orflame.Itmaytakeseveralhoursforoxygenlevelsinfabrics
toreturntonormal.