Invacare HF2E540M9 Oxygen Equipment User Manual


 
SECTION 4—INSPECTION AND COMPRESSOR INTERFACE
HomeFill
®
Cylinder 12 Part No 1163156
Checking Cylinder Pressure
ƽ WARNING
NEVER use tools of any kind to connect/disconnect the
cylinder and the compressor. Otherwise, severe injury
and/or damage may occur.
DO NOT drop oxygen cylinders. Use two hands when
handling/transporting oxygen cylinders. Otherwise, injury
or damage may occur.
DO NOT use a liquid leak detector to test for leaks.
NOTE:Forthisprocedure,refertoFIGURE4.1onpage 13.
1. Installaconserverorregulatorontothecylinderaccording
tothedevicemanufacturerinstructions.
2. Ensurecylindervalveisopen.Openthecylindervalveby
rotatingcounterclockwise.
3. Examinethegaugeontheconnectedconserveror
regulator.Performoneofthefollowing:
•Ifthecylinderpressureislessthan1500psig,fillthe
cylinder.Connectthecylindertothecompressor.Refer
toFillingtheCylinderonpage 13.
•Ifthecylinderpressureisgreaterthan1500psig.,DO
NOTattempttofillthiscylinder.Itmaynotfill.Use
cylinderasnormal.