Invacare HF2E540M9 Oxygen Equipment User Manual


 
SECTION 4—INSPECTION AND COMPRESSOR INTERFACE
Part No 1163156 11 HomeFill
®
Cylinder
External Examination
1. Examinetheoutsideofthecylinderforthefollowing
conditionsandreplacethecylinderiftheyexist:
•Dentsordings
•Arcburns
•Oilorgrease
•Anyothersignsofdamagethatmightcauseacylinder
tobeunacceptableorunsafeforuse.
2. Examinethecylinderforevidenceoffireorthermal
damage.Evidenceincludescharringorblisteringofthe
paint,orotherprotectivecoatingorheatsensitive
indicator.Iffireorthermaldamageisfound,replacethe
cylinder.
3. Inspectthecylinderfillportforthefollowingandreplace
iffound:
•Debris,oilorgrease
•Noticeablesignsofdamage
•Signsofcorrosioninsidethevalve
•Signsofexcessiveheatorfiredamage