Invacare HF2E540M9 Oxygen Equipment User Manual


 
SECTION 4—INSPECTION AND COMPRESSOR INTERFACE
Part No 1163156 13 HomeFill
®
Cylinder
Filling the Cylinder
NOTE:Forthisprocedure,refertoFIGURE4.1thruFIGURE4.6.
1. Checkthecylinderpressurebeforeattemptingtofill
cylinder.RefertoCheckingCylinderPressureonpage 12.
2. Closethecylindervalvebyrotatingclockwise(FIGURE
4.1).
3. Removetheconserverorregulatoraccordingtothedevice
manufacturerinstructions(FIGURE4.1).
FIGURE 4.1 Component Identification
4. Removethepostvalvecover(locatedonthecylinder)and
theconnectorfillportcovers(locatedonthecompressor)
(FIGURE4.2).
Conserver/
Regulator
Cylinder
Cylinder
Valve
Gauge
CONSERVER/REGULATOR
INSTALLED
CONSERVER/REGULATOR
REMOVED