Invacare FDX-MCG Mobility Aid User Manual


 
8 SETUP/MAINTENANCE
Part No 1163181
81 Invacare® FDX®
4. Remove the tie-wraps securing the right side wiring bundle together (FIGURE3, Detail “A”).
5. Remove the two mounting screws securing the controller bracket to the wheelchair base frame (FIGURE2).
6. Lifting up on the controller bracket, disengage the controller bracket from the wheelchair base frame.
7. Using one hand to support the controller bracket, use the other hand to disconnect the two motor connectors, the rear battery wiring
connector, the joystick connector and the auxiliary wiring harness connector (if equipped with power seating system, running light package or
other powered accessory) from the controller (FIGURE3, Detail “A”).
8. Set the controller bracket aside.
9. Disconnect the rear battery wiring connector from the rear battery connector (FIGURE3, Detail “B”).
10. Remove the rear battery from the wheelchair base frame.
ƽ
WARNING
When removing or installing the rear battery, DO NOT allow the controller bracket to come in contact with the top of
the battery. ALWAYS remove the controller bracket from the wheelchair base frame when removing the batteries -
otherwise injury or damage may occur.