8 SETUP/MAINTENANCE
Invacare® FDX®
76 Part No 1163181
8.4 Batteries
ƽ
WARNING
The use of rubber gloves is recommended when working with batteries.
Invacare strongly recommends that battery installation and battery replacement always be done by a qualified technician.
After ANY adjustments, repair or service and before use, make sure all attaching hardware is tightened securely -
otherwise injury or damage may result.
GP34 batteries weigh 42 pounds each. Use proper lifting techniques (lift with your legs) to avoid injury.
DO NOT lift batteries by the hook and loop battery strap. The hook and loop battery strap should be used to assist in
removing the batteries from the wheelchair base frame.
DO NOT tip the batteries. Keep the batteries in an upright position.
NEVER allow any of your tools and/or battery cable(s) to contact BOTH battery post(s) at the same time. An electrical
short may occur and serious personal injury or damage may occur.
When tightening the clamps, always use a box wrench. Pliers will “round off” the nuts. NEVER wiggle the battery
terminal(s)/post(s) when tightening. The battery may become damaged.
The POSITIVE (+) RED battery cable MUST connect to the POSITIVE (+) battery terminal(s)/post(s), otherwise serious
damage will occur to the electrical system.
Install protective caps on POSITIVE (+) and NEGATIVE (-) battery terminals.
Do not remove fuse or mounting hardware from POSITIVE (+) red battery cable mounting screw.
Unless otherwise indicated, make sure power to the wheelchair is OFF before performing these procedures.
After ANY adjustments, repair or service and before use, make sure all attaching hardware is tightened securely -
otherwise injury or damage may occur.