9 TROUBLESHOOTING
Invacare® FDX®
92 Part No 1163181
Information Gauge Display Diagnostics
DISPLAY DESCRIPTION DEFINITION COMMENTS
All LEDs are off. Power is off.
All LEDs are on. Power is on. Fewer than three LEDs on implies reduced
battery charge.
Left RED LED is flashing. Battery charge is low. The batteries should be charged as soon as
possible.
Left to Right “chase” alternat-
ing with steady display.
Joystick is in programming, inhibit
and/or charging mode.
The steady LEDs indicate the current state
of the battery charge.
All LEDs are flashing slowly. Joystick has detected Out-of-
Neutral-at-Power-Up mode.
Release the joystick back to Neutral.
Information