Invacare At'm Mobility Aid User Manual


 
SECTION 8—BATTERIES
Part No. 1125035 35 At’m
REMOVING/INSTALLING THE BATTERY PACK
ƽ WARNING
Never unplug the battery cables by pulling on the wires. Unplug the battery cables
by pulling on the connector ONLY. Otherwise injury or damage may result.
NOTE: For this procedure, refer to FIGURE 8.2.
NOTE: If there is battery acid in the bottom of the battery pack, tray or on the sides of the
battery(ies), apply baking soda to these areas to neutralize the battery acid. Before reinstalling the
existing or new battery(ies), clean the baking soda from the battery tray or battery(ies) being sure
to avoid contact with skin and eyes. Determine source of contamination. Never install/reinstall a
battery with a cracked or otherwise damaged case.
REMOVING
1. Verify the joystick on/off button is in the off position and disconnect the joystick. Refer
to Disconnecting/Connecting the Joystick on page 32.
2. Disconnect the battery connector from the wheelchair (RED and BLACK connector).
3. Using the battery handle, lift the battery pack up and away from the wheelchair base.
INSTALLING
1. Verify the joystick on/off button is in the off position and disconnect the joystick. Refer
to Disconnecting/Connecting the Joystick on page 32.
2. Position the battery pack into brackets located on the rear of the wheelchair.
3. Connect the battery connector to the connector on the rear of the wheelchair (RED and
BLACK connector).
4. Connect the joystick. Refer to Disconnecting/Connecting the Joystick on page 32.
FIGURE 8.2 - REMOVING/INSTALLING THE BATTERY PACK
Battery Pack
Handle
Battery Pack
RED and BLACK
Battery Connector
Brackets
RED and BLACK
Controller Connector
Wheelchair Base