Invacare 636959 Oxygen Equipment User Manual


 
Part No 636959 7 Aspirator Suction Unit
Replacing the Hydrophobic Filter
NOTE:Forthisprocedure,refertoFIGURE 3.
NOTE:Replacethehydrophobicfilterwheneitherofthefollowingconditionsoccur:(1)the
filterisdirtyandtheaspiratorhaslostsuction,or(2)thereisliquidinthehydrophobic
filterassemblyduetoextendeduseoroverflow.
NOTE:Thehydrophobicfiltershouldbereplacedeverytwomonths.However,ifoverflow
occurs,changethefilterimmediately.
CAUTION
Use only Invacare recommended hydrophobic filters. DO NOT replace with
other types of bacteria filters.
1. Removetheelbowconnectorendofthehydrophobicfilterassemblyfromthe
topofthecollectionjar.
2. Separatetheoppositeendofthehydrophobicfilterassemblyfromthevacuum
gauge.
3. Disassemblethehydrophobicfilterassemblybydisconnectingthetwo
connectiontubesfromthehydrophobicfilterhousing.
4. Discardthefilter.
5. Reassemblethehydrophobicfilterassemblywithanewfilter.
6. Reattachthehydrophobicfilterassemblyontothecollectionbottlelidand
vacuumgauge.
7. Ifnecessary,twisttheelbowconnectortoalignwiththecollectionjar.
FIGURE 3 Hydrophobic Filter Replacement
Hydrophobic
Filter
Collection Bottle Lid
Connection
Tubing
Connection
Tubing
Vacuum
Adjustment
Knob