Invacare 1200 Personal Lift User Manual


 
GENERAL GUIDELINES
Part No. 1150696 9 Traverse Lift/Transporter
Performing Maintenance
RefertoMAINTENANCEANDTROUBLESHOOTINGonpage 26foramaintenance
scheduleandprocedures.
Regularmaintenanceofliftsandaccessoriesisnecessarytoassureproperoperation.
Castersandaxleboltsrequireinspectionseverysixmonthstocheckfortightness,wear,
debris(suchashairanddirt)andthatthey rollfree.
Afterthefirsttwelvemonthsofoperation,inspectallglides,rollers,andfastenersfor
wear.Makethisinspectioneverysixmonthsthereafter.Ifthemetalisworn,theparts
MUSTbereplaced.
BearingscrewsontheliftcolumnMUSTbeadjustedregularly.Otherwise,theliftcolumn
mayseizeandbecomeinoperable.RefertoTroubleshootingonpage 28.
DONOTattempttoopenthemotororobtainlocalserviceasthiswillVOIDthe warranty
andmayresultindamageandacostlyrepair.ConsultyourdealerorICCIforfurther
information.
Electrical - Grounding Instructions
DONOT,underanycircumstances,cutorremovetheroundgroundingprongfromany
plug.Somedevicesareequippedwiththreeprong(grounding)plugsforprotection
againstpossibleshockhazards.Whereatwoprongwallreceptacleisencountered,itis
thepersonalresponsibilityandobligationofthecustomertocontactaqualified
electricianandhavethetwoprongreceptaclereplacedwithaproperlygroundedthree
prongwallreceptacleinaccordancewiththeNationalElectricalCode.Ifyoumustusean
extensioncord,useonlyathreewireextensioncordhavingthesameorhigherelectrical
ratingasthedevicebeingconnected.Inaddition,ICCIhasplacedRED/ORANGE
WARNINGTAGSonsomeequipment.DONOTremovethesetags.Carefullyread
battery/batterychargerinformationpriortoinstalling,servicingoroperatingyourlift.
Pinch Points
ƽ WARNING
Pinch points exist at base of lift. Be careful, injury could occur.