SECTION 7—MAINTENANCE
Roze™ Stand Up Patient Lift 24 Part No. 1150703
FIGURE 7.1 Replacing the Electric Actuator
Replacing the Leg Actuators
NOTE: For this procedure, refer to FIGURE 7.2 on page 25.
NOTE: If the legs of the patient lift do not open and close properly, it may be necessary to replace one or both of the leg actuators.
1. Test the operation of the legs:
Press the CLOSE button on the hand control to close the legs.
Press the OPEN button on the hand control to open the legs.
2. While holding on to the mast handles, carefully tip the patient lift back until it rests on the floor.
NOTE: It is necessary to remove both small base covers before removing the large base cover.
3. Remove the sixteen screws that attach the small base covers to the base.
4. Remove the eight pan head screws that attach the large base cover to the base.
NOTE: If the left or the right leg of the patient lift does not operate properly, perform the following steps on the side of the base necessary
to replace the actuator. If both legs do not operate properly, perform this replacement procedure on both sides of the base.
5. Locate and disconnect the leg actuator lead wires.
Cap
Screw
Washer
Actuator Cap
Bushing
Cap
Flat
Washer
Bearing
Screw
Cap
Bearing
Washer
Electric Actuator
Bushing
Washer
Bearing
Nut
Cap
Lift Arm Actuator
Mounting Bracket
Actuator Lead
Routing Hole
TOP
BOTTOM
Washer
Nut
Flat Washer
Nylon
Washers
Mast Assembly
Actuator Mounting
Bracket
Nylon
Washers
Bearing