Invacare 1150703 Personal Lift User Manual


 
SECTION 3—OPERATION
Part No. 1150703 13 Roze™ Stand Up Patient Lift
Activating a Mechanical Emergency
Release
Primary Emergency Release
NOTE: For this procedure, refer to FIGURE 3.2.
NOTE: This procedure will bring the boom down or lift it up if the hand
control is not functioning properly.
To activate the primary emergency release, insert a pen into the
hole labeled Emergency Up or Emergency Down on the control
box.
Secondary Emergency Release
NOTE: For this procedure, refer to FIGURE 3.3.
NOTE: All patient lift actuators are equipped with a mechanical
emergency release. The mechanical release will enable the actuator to
retract without power.
NOTE: Use the primary emergency release first before using the
secondary emergency release procedure. This procedure should only be
used if the primary emergency release procedure is not functioning or is
unreachable.
NOTE: The lift MUST be under a load for the mechanical release to
function.
To activate the secondary emergency release, pull up on the RED
emergency grip and pull down on the boom at the same time.
Charging the Battery
NOTE: For this procedure, refer to FIGURE 3.4.
NOTE: Invacare recommends the battery be recharged daily to prolong
battery life.
NOTE: An audible alarm will sound when battery is low.
1. Lift UP on the handle on the back of the battery.
2. Lift the battery up and away from the control box.
ƽ CAUTION
Make sure there is an audible click when mounting
battery on the battery charger. Otherwise, injury
or damage may occur.
3. Place the battery on the battery charger. Push the top of the
battery against the mounting bracket until there is an
audible click.
NOTE: The charge LED will illuminate. When charged, the LED will
stop illuminating.
NOTE: It will take approximately four hours to charge a battery that
requires a full charge.
4. Lift UP on the handle on the back of the battery.
5. Lift the battery up and away from the battery charger.
ƽ CAUTION
Make sure there is an audible click when mount-
ing battery on the battery charger. Otherwise,
injury or damage may occur.
6. Place the battery on the control box. Push the top of the
battery against the mounting bracket until there is an
audible click.
FIGURE 3.2 Primary Emergency Release
FIGURE 3.3 Primary Emergency Release
FIGURE 3.4 Primary Emergency Release
Emergency Up
Hole
Emergency Down
Hole
Control Box
RED Emergency
Grip
Battery Charger
(STEP 3)
Control Box
(STEP 6)
An audible “click” will be heard
when properly installed
(STEPS 3 and 6)
Handle
(STEPS 1, 2, 4 and 5)
Battery