SYMPTOMS
POSSIBLE
CAUSES
CORRECTION
Unit does not turn
on when the
START/STOP
button is pushed.
Batteries have
run down.
Replace them
with two new AAA
alkaline batteries.
Battery polarities
have been posi-
tioned incorrectly.
Re-insert the
batteries in the
correct positions.
EE measurement
error symbol shown
on display or the
blood pressure
value is displayed
excessively low
(or high).
Cuff has been
placed incorrectly.
Wrap the cuff prop-
erly so that it is
positioned correctly.
Did you talk or
move during
measurement?
Measure again.
Keep wrist
steady during
measurement.
Shaking of the wrist
with the cuff on.
E1 error symbol
shown on display
Air circuit
abnormality.
Measure again.
E2 error symbol
shown on display
Inflation pressure
exceeding 300 mmHg.
Switch the unit off,
then measure again.
E3 error symbol
shown on display
Data error. Remove and reload
the batteries.
Power Source: Two 3V DC (AAA) Alkaline batteries
Measurement
Method:
Oscillometric
Measurement
Range:
Pressure: 40~250 millimeters Mercury
(mmHg)
Pulse: 40~199 beats/minute
Accuracy:
Pressure: ±3 mmHg
Pulse: ±5% of reading
Pressure Sensor: Semi-conductor
Inflation: Pump driven
Deflation: Automatic Pressure release valve
Memory Capacity: 60 memories for each User (120 total)
Auto-shut-off: 1 minute after last button operation
Operation
Environment:
Temperature: 50°F~104°F (10°C~40°C)
Humidity: 40%~85% RH max.
Storage
Environment:
Temperature: 14°F~140°F (-10°C~60°C)
Humidity: 10%~90% RH max.
Weight: 0.30 lbs (without batteries)
Wrist
Circumference:
5.3”~8.7” (13.5~22cm)
Dimensions:
2.6”(L) x 1.3”(W) x 3.1”(H)
67mm(L) x 32mm(W) x 78mm(H)
Type BF
Device and cuff are designed to
provide special protection against
electrical shocks.
SPECIFICATIONS
Note: These specifications are subject to change
without notice.
TROUBLESHOOTING
If any abnormality arises during use, please check the following points:
Note: If the unit still does not work, contact HoMedics Consumer
Relations. Under no circumstance should you disassem-
ble or attempt to repair the unit by yourself. Contact infor-
mation for HoMedics Consumer Relations Department can
be found on the warranty page (See page 35).
BPW-101IB_ENGSP.qxd 10/24/05 11:14 AM Page 33