Health O Meter 2000KL Scale User Manual


 
25
CALIBRATION PATH
The calibration is performed in Kg or Lb, according to the units of measure used upon entering
into the calibration path.
Operator Action Display
1. Press ON/OFF button and immediately press
and hold HOLD/RELEASE button for 3-4
seconds
Calibration
Process
2. Press ENTER Enter load weight
400.0
3. Using cd keys and/or the keypad, set the
calibration load to at least 200 lbs (for best
results and greater accuracy, use a 400 lb
weight). Press ENTER
Zero calibration
Please clear the scale
4. Please clear the stretcher and press ENTER Zero calibration
Please wait
5. Please wait without touching or shaking the
scale until the zero calibration process is
finished
Weight calibration
Put: XXX.X
6. Load the stretcher with the required weight and
press ENTER
Weight calibration
Please wait
7. Please wait without touching or shaking the
scale until the calibration process is finished
Calibration
factor: X.XXXXX
8. Press ENTER Calibration
Please clear the scale
9. Please remove weight from the stretcher Rebooting
Please wait
10. Please wait until the scale resumes normal
operation
Health O Meter