MAINTENANCE SCHEDULE
Info: Only qualified persons should service and repair your Lumex patient lift. Regular
maintenance of your patient lift is necessary to ensure continuing proper and safe
operation. Read and observe the following recommended maintenance schedule:
Item Inspect
initially
Inspect/adjust
daily monthly every 6
months
Base and Casters
Ensurethatbaseislevel(allfourwheelstouchtheoor) X X X
Ensurethatcastermountingboltsandaxleboltsaretight X X
Ensurethatcastersswivelandrollfreely X X
Applyalightgreasetocasterballbearings X
Base Adjustments
Ensurethatbaseopensandclosessmoothly X X X
Ensurethatbaselockssecurelyinthreepositions X X
Ensurethatlockingknobisinplaceandoperatingproperly X X
Inspectallbasehardwareforwear X
Mast
Ensurethatmastisfullyengagedandlockedintomastsleeve X X
Inspectmastforbendsordeections X
Inspectallmasthardwareforwear X X
Boom
Inspectboomforbendsordeections X X
Ensurethatboomiscenteredbetweenthebaselegs X X X
Inspectallboomhardwareforwear X
Spreader Bar
Inspecthangerpostforwearordamage X
Inspectslinghookforwearordeections X X
Inspectallspreaderbarhardwareforwear X X
Slings
Inspectslingsandstrapsforwearordamage X
Inspectmetal/plasticpartsforwearordamage X
Lubrication with light oil
Lubricateallpivotpoints X
Lubricatetopandbottomofactuator/motormounting X
Cleaning
Cleanwithmildnon-abrasivedetergent X
Pump (LF1031 only)
Ensurethatpumpmountinghardwareistight X X
Inspectpumpmountinghardwareforwear X
Electronics and actuator / motor (LF1041 only)
Ensurethatbatteryisfully
charged X X
Inspectforproperup/downoperation X
X
Ensurethatmountinghardwareistight X X
Inspectmountinghardwareforwear X