Golden Technologies GP160 Wheelchair User Manual


 
LiteRider PTC_Owner’s Manual Rev A_121911
VII. OPERATION
Battery Gauge Flashes Rapidly (even with the joystick released) See figure 15.
The control system safety circuits have operated and the control system has been prevented
from moving the wheelchair. This indicates a system trip, i.e. the VR2 has detected a
problem somewhere in the wheelchair’s electrical system.
Please follow this procedure:
Place the joystick in the center position and turn the control system off and on
again to clear the fault. If the fault does not clear, go to the next step.
Switch off the control system.
Disconnect and reconnect all connectors on the wheelchair and the control system
to make sure they are mated securely.
Check the condition of the battery.
If you can’t find the problem, go to the “Self-Help Guide” section below.
Switch on the control system again and try to drive the wheelchair. If the safety
circuits operate again, switch off and do not try to use the wheelchair. Contact
your Golden Technologies representative for service.
Self-Help Guide
If a system trip occurs, you can find out what has happened by counting the
number of LEDs on the battery gauge that are flashing. See page 25.


24
Figure
1
5
Battery Charge
LED Array
On
/
Off
Button
Horn Button
Speed Setting
FASTER
Speed Setting
LED Array
Speed Setting
SLOWER