The HEPA-filtered exhaust air can be recirculated into the
laboratory or exhausted outside with the addition of the exhaust
connection kit and remote blower listed in Chapter 7.
The XPert Filtered Balance Station does not include the motorized
impeller and must be exhausted to the outside by a remote blower.
The XPert Filtered Balance Station remote blower may be
switched from the blower switch on the enclosure. See electrical
Requirements in Chapter 2, and Electrical Supply Connections in
Chapter 3. Only one 6" exhaust connection is required for the 2',
3', or 4' XPert Filtered Balance Stations to exhaust to the outside.
Data for the exhaust volume, noise pressure and enclosure static
pressure loss are listed for each filtered enclosure model at face
velocities of 60, 75, 80, 90, 100, and 105 fpm.
Enclosure
Width Model Description
Face
Velocity
(fpm)
Exhaust
Volume
(CFM)
XPert
Station
Initial Static
Pressure
Loss with
HEPA filter
(in w.g.)
XPert
System
Noise
Pressure
db(A)
Xpert
System
Max.
Equivalent
Resistance
of 6" Duct
(Ft)
Xpert
System
Max.
External
Static
Pressure
2' 2' XPert Filtered Balance System
2' XPert Filtered Balance Station
60
75
80
90
100
105
85
110
115
130
145
155
.35"- .41"
.45" - .54"
.48" - .57"
.56" - .65"
.62" - .72"
.65" - .76"
48-53
49-55
50-56
51-57
53-58
54-59
400
220
200
140
80
70
.20"
.19"
.18"
.16"
.13"
.12"
3' 3' XPert Filtered Balance System
3' XPert Filtered Balance Station
60
75
80
90
100
105
130
165
175
200
220
230
.37" - .43"
.50" - .58"
.53" - .62"
.61" - .71"
.68" - .79"
.71" - .88"
48-53
52-56
53-57
55-58
58-61
60-64
130
65
50
35
25
20
.15"
.12"
.11"
.09"
.08"
.07"
4' 4' XPert Filtered Balance System
4' XPert Filtered Balance Station
60
75
80
90
100
105
175
220
235
265
295
310
.40" - .46"
.53" - .62"
.57" - .66"
.68" - .78"
.76" - .87"
.80" - .91"
49-54
54-58
58-61
60-63
62-66
64-67
100
60
50
35
25
20
.20"
.19"
.18"
.16"
.13"
.12"
XPert
Station