Chapter 6: Maintaining Your Filtered Enclosure
To start the calibration, use the Labconco 1000 Manual and enter
the “CALIBRATIION” mode on the display. Follow the
instructions on the display and alter the low and high exhaust
volumes with the speed control. The following suggested inflow
face velocity speeds are recommended to successfully calibrate.
Typical low air alarms are set 10-30 fpm below operational speeds.
Typical yellow cautions are set 3-10 fpm above the red low air
alarm. Be sure to enter “Done” after completing the calibration in
“SETUP.” Once completed, go to “RUN” and hit “ENTER.”
Low Air Alarm
Set Point
Nominal Enclosure
Operating In Flow
Speed
Low
Calibration
Set Point
High
Calibration
Set Point
50 fpm 60 fpm 50-60 fpm 100-120 fpm
60-70 fpm 80 fpm 50-70 fpm 100-140 fpm
70-90 fpm 100 fpm 60-80 fpm 110-140 fpm
Factory Settings
60 ± 10 fpm 90 ± 10 fpm 60 ± 10 fpm 90 ± 10 fpm