Fluke 561 Thermometer User Manual


 
Infrared Thermometer
Operating the Thermometer
7
D:S=12:1
38 mm @
300 mm
75 mm @
900 mm
132 mm @
1500 mm
1.5 " @
12 "
3 " @
36 "
5.3 " @
60 "
S
D
efh005f.eps
Figure 5. Distance and Spot Size
Field of View
Make sure that the target is larger than the spot size. The smaller the target, the closer you should be to it.
See Figure 6.
Ye s
No
efh004f.eps
Figure 6. Field of View
Emissivity
Emissivity describes the energy-emitting characteristics of materials. Most organic materials and painted or
oxidized surfaces have an emissivity of about 0.95.
If possible, to compensate for inaccurate readings that may result from measuring shiny metal surfaces,
cover the surface to be measured with masking tape or flat black paint (< 148 °C/300 °F) and use the high
emissivity setting. Allow time for the tape or paint to reach the same temperature as the surface beneath it.
Measure the temperature of the tape or painted surface.
If you cannot paint or use tape, then you can improve the accuracy of your measurements with the
emissivity selector. Even with the emissivity selector, it can be difficult to get a completely accurate infrared
measurement of a target with a shiny or metallic surface. Experimentation, using the probe to determine
benchmark temperatures, and experience will help you choose the best setting for specific measurements.
The Thermometer has three emissivity settings: low (0.3), medium (0.7), and high (0.95). Refer to Table 2.
The reference to emissivity settings in the table are suggestions for typical situations. Your particular
situation may differ.