Fluke 9105 Thermometer User Manual


 
9.1.3 Serial Operation
Once the cable has been attached and the interface set up properly the control
-
ler will immediately begin transmitting temperature readings at the pro
-
grammed rate. The serial communications uses 8 data bits, one stop bit, and no
parity. The set-point and other commands may be sent via the serial interface to
set the temperature set-point and view or program the various parameters. The
interface commands are discussed in section . All commands are ASCII charac
-
ter strings terminated with a carriage-return character (CR, ASCII 13).
9.2 IEEE-488 Communication
The IEEE-488 interface is available as an option. Instruments supplied with this
option may be connected to a GPIB type communication bus which allows
many instruments to be connected and controlled simultaneously. To eliminate
noise, the GPIB cable should be shielded
9.2.1 Setup
To use the IEEE-488 interface first connect an IEEE-488 standard cable to the
back of the calibrator. Next set the device address. This parameter is pro-
grammed within the IEEE-488 interface menu. The IEEE-488 interface param-
eters menu is outlined in Figure 6.
To enter the IEEE-488 parameter programming menu first press “EXIT” while
pressing “SET” and release to enter the secondary menu. Press “SET” repeat-
edly until the display reaches “PrObE”. This is the menu selection. Press “UP”
repeatedly until the IEEE-488 interface menu is indicated with “IEEE”. Press
“SET” to enter the IEEE-488 parameter menu. The IEEE-488 menu contains
the IEEE-488 address parameter.
9.2.1.1 IEEE-488 Interface Address
The IEEE-488 address is prompted with “AddrESS”. Press “SET” to program
the address. The default address is 22. Change the device address of the calibra
-
tor if necessary to match the address used by the communication equipment by
pressing “UP” or “DOWN” and then “SET”.
9.2.2 IEEE-488 Operation
Commands may now be sent via the IEEE-488 interface to read or set the tem
-
perature or access other controller functions. All commands are ASCII charac
-
ter strings and are terminated with a carriage-return (CR, ASCII 13). Interface
commands are listed below.
9105/9107 45
9 Digital Communication Interface