Fluke 1521 Thermometer User Manual


 
4. Verify the accuracy at 0Ω,4kΩ,10kΩ,40kΩ, 100 kΩ, and 500 kΩ.
5. Set the meter calibration date to the present date (see Section 7.8.18).
6. Set the meter due date to one year from the present date (see Section 7.8.19).
1521 Handheld Thermometer Readout
User’s Guide
44