Fluke 1521 Thermometer User Manual


 
Description Command Example
Response
Format
Response
Example Range
Set resolution digits re[solution]=n res=1 1 to 3
Read time ti[me] time ti: hh:mm:ss ti: 16:23:45
Set time
Note: You must set hours, min
-
utes, and seconds for the time to
be set correctly.
ti[me]=hh:mm:s
s
ti=08:15:0
0
00:00:00 to
23:59:59
Read identification information *i[dn?] *idn? <manufacturer>,
<model>,
<serial number>,
<firmware version>
Hart,
1521,
95001,
2.11
Read model and firmware version *v[ersion] *ver ver.<model>,
<firmware version>
ver.1521,
2.11
Legend: [] Optional Command data
{} Returns either information
n Numeric data supplied by user
9 Numeric data returned to user
x Character data returned to user
8.2.2 Calibration Commands
The following group of commands directly or indirectly affect the accuracy of
the instrument. These commands should only be used by qualified, authorized
personnel. Calibration parameters can only be changed if protection is dis-
abled by entering the proper pass-code using the command *pas= ddddd.
Entering an invalid passcode or cycling the power re-establishes protection of
the calibration parameters. The parameters can be read without disabling the
passcode protection.
Description Command Example
Response
Format
Response
Example Range
Unprotect calibration parame
-
ters with the pass-code
*pas[scode]=nnnnn *pas=12345
Read date da[te] date da: yyyy-mm-dd da: 1999-03-28
Read passcode protection
state
*pas *pas ps:9 ps:0 0 or 1
Set date da[te]=yyyy-mm-dd da=1999-03-28 1999-01-01 to
2100-12-31
Read probe serial number se[rial] ser ser: xxxxxxxxxx ser: 100231A
Set probe serial number se[rial]=xxxxxxxxxx ser=001 0 to 9, A to Z,
., -, /
Read probe type p[robe] pr pr: {ITS-90, IEC751,
CVD, YSI400, THERM}
pr: ITS-90
39
8 Communications Interface
Communication Command List