Emerson 3051S Pacemaker User Manual


 
Quick Installation Guide
00825-0100-4801, Rev HA
February 2008
Rosemount 3051S
6
STEP 2: CONSIDER HOUSING ROTATION
STEP 3: SET SWITCHES AND JUMPERS
If alarm and security adjustment option is not installed, the transmitter
will operate normally with the default alarm condition alarm high and
the security off.
To improve field access to wiring or to
better view the optional LCD display:
Figure 1. Transmitter Housing
Set Screw
1. Loosen the housing rotation set
screw.
2. First rotate the housing clockwise
to the desired location. If the
desired location cannot be
achieved due to thread limit,
rotate the housing counter
clockwise to the desired location
(up to 360° from thread limit).
3. Retighten the housing rotation set
screw.
PlantWeb Junction Box
Figure 2. Transmitter Switch and Jumper Configuration
PlantWeb Junction Box
Slide the security and alarm
switches into the preferred position
by using a small screwdriver.
(A LCD display or an adjustment
module must be in place to activate
the switches.)
Pull the jumpers out and rotate 90°
into desired position to set the security
and alarm.
Housing Rotation Set
Screw (3/32-inch)
Alarm
Security
Meter/
Adjustment
Module
Security
Alarm
4801_QIG_RevHA.fm Page 6 Tuesday, February 26, 2008 12:17 PM