Quick Installation Guide
00825-0100-4593, Rev BA
July 2009
Rosemount 1151
16
Canadian Certifications
Canadian Standards Association (CSA) Approvals
E6 Explosion Proof for Class I, Division 1, Groups C and D; Class II, Division 1, Groups E,
F, and G; Class III, Division 1 Hazardous Locations. Suitable for Class I, Division 2,
Groups A, B, C, and D; CSA enclosure type 4X. Factory Sealed.
I6 Intrinsically safe for Class I, Division 1, Groups A, B, C, and D hazardous locations
when connected per Drawing 01151-2575. For entity parameters see control drawing
01151-2575. Temperature Code T2D.
Measurement Canada Approvals
C5 Accuracy Approval to the Electricity and Gas Inspection Act for the purchase and sale
of natural gas.
European Certifications
E8 ATEX Flameproof
Certification Number CESI03ATEX037
ATEX Marking II 1/2 G
EEx d IIC T6 (–40 °C T
amb
40 °C)
EEx d IIC T4 (–40 °C T
amb
80 °C)
1180
V = 60 Vdc maximum
I1 ATEX Intrinsic Safety and Combustible Dust Certification
Certification Number BAS99ATEX1294X
ATEX Marking II 1 GD
Ex ia IIC T5 (–60 °C T
amb
40 °C)
Ex ia IIC T4 (–60 °C T
amb
80 °C)
Dust Rating: T90 °C (T
amb
= –20 °C to 40 °C)
IP66
1180
Entity Parameters
Special Conditions for safe use (x):
The apparatus is not capable of withstanding the 500V insulation test required by
EN60079-1. This must be taken into account when installing the apparatus.
U
i
= 30V
I
i
= 125 mA
P
i
= 1.0 W (T4) or 0.67 W (T5)
C
i
= 0.034 F
L
i
= 20 H
4593_Rev_BA.fm Page 16 Friday, August 7, 2009 3:38 PM