Braun L200UARS Wheelchair User Manual


 
Page 42
Replace missing, worn
or illegible decals.
Failure to do so may
result in serious bodily
injury and/or property
damage.
W
A
RNING
Decals and Antiskid
Decals
See Decals on page 41 also.
Note: Clean surfaces as
detailed on opposite page
before posting decals.
25675
W
A
RNING
Lift installation and servicing prohibited by
anyone who has not been certified by The Braun
Corporation Sales and Service School. Certified
service technicians should call 1-800-THE LIFT
to receive applicable installation/service manual.
Failure to follow this policy may result in serious
bodily injury and/or property damage.
25675
15707
W
A
RNING
Discard flanged platform lock nuts if
loosened or removed. Do not reuse!
Replace with nuts provided in lift
installation parts package. Failure to
do so may result in serious bodily
injury and/or property damage. See
owner's manual for instructions.
15707
W
A
RNING
Contact The Braun
Corporation before
adjusting hydraulic
pressure relief
valve. Failure to
do so may result
in serious bodily
injury and/or
property damage.
22249
1786625652
16735
W
A
RNING
Do not use this hole for lift securement. Failure
to follow this instruction may result in serious
bodily injury and/or property damage.
16735
24668
17309
The Braun Corporation certifies that
this wheelchair lift conforms to all
applicable requirements of the National
Standards for School Buses in effect at
the time of its manufacture.
24668
22249
Fairfield, NJ USA
Oslo, Norway
Clearwater, FL USA
Kitchener, Ontario Can.
Huntington Beach, CA USA
Rocklea, Queensland Aust.
"Providing Access to the World"
International Corporate Hdqrs: P.O. Box 310 Winamac, IN 46996 USA
1-800-THE LIFT (219) 946-6153 FAX: (219) 946-4670
12375
®
Parking Courtesy
Requested
12375
25652
W
A
RNING
Support platform
before pulling
quick-release pin.
25948
25948
BLU BLU
BLACK
BLACK
BLACK
BLACK
Interlock
Jumper
Ground Ground
Up
Solenoid
Up
Solenoid
NO
NC
COM
INTERLOCK SWITCH
Previously
Cut
Jumper
Distance to Interlock Switch 5 feet or less
NC
NO
COM
Distance to Interlock Switch more than 5 feet
* Use 10 gauge wires
(Black preferred)
Lift Interlock and Instructions
located behind this panel
NOTICE
Connector
To solenoid
ground stud
Previously
Cut
Jumper
To solenoid
ground stud
Connector
Connector
• 15 AMP rating
• customer supplied
INTERLOCK SWITCH
• 15 AMP rating
• customer supplied
CHASSIS
GROUND
* Use 12 gauge wires
(Black preferred)
Interlock Installation Instructions
1. Note: Terminal connected to UP solenoid mounting
stud provides ground source for lift controls.
2. To connect remote interlock switch:
a. Remove (disconnect) jumper male plug.
b. Snap plastic female socket into pre-
punched top housing plate.
c. Cut jumper wire and connect remote
ground source.
d. Reconnect male plug and female socket.
See Specifications below.
3. With lift unoccupied,
test all lift functions
to ensure proper
interlock installation
(with interlock switch
opened and closed).
Specifications
Failure to use proper wire gauges may cause slow FOLD
operation.
17309
Note:
See Specifications
below.
d. Reconnect male plug
and female socket.
c. Cut jumper wire and
connect remote
ground source.