A&D FS-15Ki Scale User Manual


 
16
5. Press the LO key. Then, the last LO % weight
entered will flash and the left part of analog display
turns on.
6. Enter the LO limit weight via the 10-key pad.
If you hit the wrong key, press the C key and the
LO limit % is re-displayed. Continue data entry again.
You can not enter the “+” or “-” sign for the LO limit.
7. Press the ENT key to store the LO limit % in memory.
8. The display shows 5tored for a few seconds and
returns to weighing mode.
The HI/LO limit % has 2 decimal places fixed in the setting mode above.
The HI/LO limit weight entered in the previous section will not be converted to %,
and vice versa.
Example:
Target weight: 3.000 kg, HI limit weight: 1.00 %, LO limit weight: 0.50 %
Then,
Upper limit = 3.000 kg + 3.000kg x 1.00 % = 3.030 kg
Lower limit = 3.000 kg - 3.000kg x 0.50 % = 2.985 kg
8-2. Upper & Lower limits weight setting mode
The UPPER & LOWER LIMITS WEIGHT SETTING MODE uses an upper limit weight
and a lower limit weight. The target weight is not used. These upper (HI) and lower
(LO) limits are entered via the 10-key pad directly.
Set the Function “F7-0 Upper & Lower limit weight setting”.
The formula to compare is as follows:
LO < Lower limit value OK Upper limit value < HI
Setting the upper (HI) limit weight and lower (LO) limit weight
Set the Function “F7-0 Upper & Lower limit weight setting”.
1. Press the HI key and the last HI limit weight
entered will flash and the right part of analog display
turns on.
2. Enter the HI limit weight via the 10-key pad.
If you hit the wrong key, press the C key and the old
HI limit weight is re-displayed. Continue data entry again.
You can enter minus value using the ± key. This
key toggles the sign.
-0000.60
-5tored
-0000.50
-002.340
-003.050