A SERVICE OF

logo

11
Note If the meter does not detect the temperature probe, it will
switch automatically to manual temperature compensation
with the temperature adjustable through the up and down
arrow keys. The “°C” symbol will blink on the LCD.
Note All external cables to be connected to the rear panel should
be ended with cable lugs.
Analog output: Connect an external recorder with a 2-wire
cable to these terminals (#1 on page 7) paying attention
to the correct polarity. A wide variety of output signals,
either in V or in mA, is available to fit most standards.
Contact 1 and 2: Connect the dosing devices to these
terminals (#4 and #7 on page 7) in order to activate and
deactivate them according to the selected control param-
eters.
mA Input: to switch to mA input signal from a
conductivity transmitter (e.g. HI8936, HI98143
or HI98144 series) see setup procedure (code
6). Connect the two signal wires from the trans-
mitter to terminals #11 on page 7, paying at-
tention to the correct polarity.
An unregulated 10 ÷ 30 VDC - 50 mA max.
power supply output (#10 on page 7) is pro-
vided to power the transmitter, if needed.
Once the installation is completed, select the appropriate
working range, the reference temperature (20 or 25°C) and
perform conductivity or TDS calibration as described in this
instruction manual. Set the control parameters according to
the process.