Use the or to make the
HI 931002 or multimeter out-
put correspond with the meter’s
value shown on the second-
ary display (e.g. 4).
Wait for approximately 30 seconds (until the reading of the
calibrator is stable).
Press CFM to confirm. The meter will switch to the second
calibration point. Repeat the above procedure.
After the desired readings are obtained, press CFM and
the meter will skip back to normal operational mode.
Note When adjusting values using the or keys it is important
to allow for sufficient response time (up to 30 seconds)
The table below lists the values of output codes along with
the calibration point values (which are the analog output
minimum and the analog output maximum) as indicated on
the display.
The secondary display indicates the current calibration point
value, while primary display indicates the current calibration
type.
OUTPUT CALIBRATION CALIBRATION CALIBRATION
TYPE CODE POINT 1 POINT 2
0-1 mA 0 0 mA 1 mA
0-20 mA 1 0 mA 20 mA
4-20 mA 2 4 mA 20 mA
0-5 VDC 3 0 VDC 5 VDC
1-5 VDC 4 1 VDC 5 VDC
0-10 VDC 5 0 VDC 10 VDC