Step 2. Check the CPU signals to the Display assembly.
Set the power switch to on ( l ).
Using a logic probe, check that the following TTL signals are toggling.
Test Location Signal Name In/Out
A7 P2(9) VSYNCEL A7 Out
A7 P2(11) HSYNCELn A7 Out
A7 P2(13) VCLK A7 Out
A7 P2(15) VID A7 Out
If the signals are incorrect, do the following:
Set the power switch to off ( O ).
Disconnect the display cable from A7 P2.
Set the power switch to on ( l ).
Check the signals again.
If the signals are now correct, the A101 Display assembly is probably faulty.