A SERVICE OF

logo

To troubleshoot display failures
Use this test to isolate display failures to the A101 Display assembly, A102 DC-DC
Converter assembly, or A7 CPU assembly.
Step 1. Check the DC-DC Converter assembly.
Set the power switch to off ( O ).
Connect the voltmeter to A102 TP1.
Set the power switch to on ( l ).
Check that the voltage reads is 210 10 Vdc.
Caution The Display assembly will be damaged if the voltage is at or above 235 Vdc.
If the voltage is incorrect, the A102 DC-DC Converter assembly is probably
faulty.
Troubleshooting the Analyzer Agilent 35670A
To troubleshoot display failures
4-22