Amplitude Accuracy
Measurement Lower Limit
(dBVrms)
Upper Limit
(dBVrms)
Measured Value
(dBVrms)
Pass/Fail
−51 dBVrms, Ch 1 −51.15 −50.85
−51 dBVrms, Ch 2 −51.15 −50.85
−43 dBVrms, Ch 1 −43.15 −42.85
−43 dBVrms, Ch 2 −43.15 −42.85
−35 dBVrms, Ch 1 −35.15 −34.85
−35 dBVrms, Ch 2 −35.15 −34.85
−27 dBVrms, Ch 1 −27.15 −26.85
−27 dBVrms, Ch 2 −27.15 −26.85
−11 dBVrms, Ch 1 −11.15 −10.85
−11 dBVrms, Ch 2 −11.15 −10.85
1 dBVrms, Ch 1
0.85 1.15
1 dBVrms, Ch 2
0.85 1.15
9 dBVrms, Ch 1
8.85 9.15
9 dBVrms, Ch 2
8.85 9.15
19 dBVrms, Ch 1
18.85 19.15
19 dBVrms, Ch 2
18.85 19.15
27 dBVrms, Ch 1
26.85 27.15
27 dBVrms, Ch 2
26.85 27.15
Serial Number:_______________________Report Number:____________________
Test Date:___/___/___
Verifying Specifications Agilent 35670A
Operation Verification Test Record - Two Channel
6of10