Toshiba LF410/LF600 Oxygen Equipment User Manual


 
LF410/LF600 LF410/LF602
3
totalizer.
Fixed-value outputs — Outputs fixed-values for
current and pulse outputs.
Zero adjustment — Executes zero adjustment
(on-stream at zero flow rate).
Output signals
Current output:
4–20mAdc (load resistance 0 to 750Ω)
Note: The current output cannot be used with
the PROFIBUS-PA ccommunication.
Digital outputs — One point (std.) and one more
point is optionally available as follows.
Digital output DO1 (std.):
Output type: Transistor open collector
Number of outputs: One point
Output capacity: 30Vdc, 200mA maximum
Digital output DO2 (opt.):
Output type: Solidstate relay output (non
polarity)
Number of outputs: One point
Output capacity: 150Vdc, 150mA maximum
or 150 V ac (peak to peak), 100mA maximum
DO1 and DO2 functions — One of the following
functions can be assigned to DO1 (std.) and/or
DO2 (opt.)
Pulse output (available only for DO1,DO2)
Pulse rate: 3.6 to 36,000,000 pulses/hr (DO1)
3.6 to 360,000 pulses/hr (DO2)
(Over 3,600,000 pulses/hr, auto-setting)
Pulse width: 0.5 to 500ms (but less than half of
the period for 100% flow rate)
Note: The same and simultaneous pulse is not
available between DO1 and DO2.)
Multi-range selection outputs (Note 1)
High, High high, Low, and/or Low low alarm
outputs (Note 2)
Empty pipe alarm output
Digital Output Active Status (DO1 and DO2)
(Note 2)
Preset count output
Converter failure alarm output
Note 1: Two outputs (DO1 and DO2) are needed for
4-range switching and forward/reverse
2-range switching.
Note 2: Normal Open (default set) or Normal Close
is selected for alarm outputs when
programming.
The Status when power failure is kept to
Normal Open.
Communications output
HART(std.)
Digital signal is superimposed on 4–20mAdc
current signal as follows:
Conforms to HART protocol
Load resistance: 240 to 750Ω
Load capacitance: 0.25µF maximum
Load inductance: 4mH maximum
PROFIBUS (opt.)
Protocol : PROFIBUS-PA
Baurate : 31.25kbps
Bus voltage : 9-30Vdc
Consumption electric current of bus:less than
16mA
Manufacture Ident-No. : 093B
HEX
Standard Ident-No. : 9740
HEX
Slave address : 0-126 (Default address is 126)
Profile : Profile Ver.3.01 for Process Control
Devices
Function blocks : AI(Flow)×1 , Totalizer×1
LCD display: Full dot-matrix 128×128 dot LCD
display (back-light provided)
The data on the LCD inside the converter can
rotate to 90, 180, and 270 degrees by a software,
without rotating the indicator itself. (Combined
type only)
Parameter settings — Parameters can be set as
follows:
• IR Switches: Three key switches are provided to
set configuration parameters.
• Digital communication: The AF900 hand-held
terminal or PROFIBUS is needed to set
parameters.
•Zero adjustment: Zero point adjustment can be
started by pressing the switch in the converter.
Damping: 0.5 to 60 seconds (selectable in 1 second
increments)
Zero and span calibration: Built-in calibration
signal source allows converter unit check.
Conditions when power fails: The outputs and
display will remain as follows when power fails.
Parameter setting values are stored in non-volatile
memory and the values will be restored when the
power returns to normal condition.
• Current output: 0mAdc
• Digital output: OFF
• LCD display: No display
• PROFIBUS: No communication
Power supply: One of the following can be selected:
• 100 to 240Vac, 50/60Hz (std.)
(allowable voltage 80 to 264Vac)
• 24Vdc (allowable voltage 18 to 36Vdc)
• 110Vdc (allowable voltage 90 to 130Vdc)
Surge protection: Arresters are installed in the
power supply, and current signal output circuit.
Case: Aluminum alloy (equivalent of IP 67)
Coating: Acrylic resin-baked coating, pearl-gray
colored