Teledyne 3000ZA-3X Oxygen Equipment User Manual


 
Trace Oxygen Analyzer Operation
Teledyne Analytical Instruments 37
4.4.1 Span Sensor
-----------------------
->Span Sensor: High
-----------------------
Pressing ENTER causes the selection after 'Sensor:' to start flashing
(Low / Mid / High). Then pressing UP or DOWN selects a different
sensor. Note that the span value directly below will change according to
which sensor is currently selected.
The high sensor covers ranges from 2 to 95% O
2
. Span gas that is
near the top of this scale should be used to calibrate this sensor.
The medium sensor covers ranges from 200 ppm to 2% O
2
. Span
gas that is near the top of this scale should be used to calibrate this
sensor.
The low sensor covers ranges from 0 to 100 ppm O
2
. Span gas that
is near the top of this scale should be used to calibrate this sensor.
4.4.2 Span Value
-----------------------
->Span 85%
-----------------------
Pressing ENTER causes the number after 'Span' to start flashing.
The number has a range of 1 ppm to 100% and should be set equal to the
concentration of the span gas used. This number corresponds to the
sensor selected above. Pressing UP or DOWN adjusts the number.
4.4.3 Span Timer
-----------------------
->Span Timer: 7 min.
-----------------------
This function sets the hold time in span mode before the analyzer
actually performs the span. Adjust to give enough time to purge with
span gas.