Teledyne 3000ZA-3X Oxygen Equipment User Manual


 
Trace Oxygen Analyzer Installation
Teledyne Analytical Instruments 25
Table 3-5: Commands via RS-232 Input
Command Description
as<enter> Immediately starts an autospan.
az<enter> Immediately starts an autozero.
st<enter> Toggling input. Stops/Starts any
status message output from the
RS-232, until st<enter> is sent
again.
The RS-232 protocol allows some flexibility in its implementation.
Table 3-6 lists certain RS-232 values that are required by the 3000ZA
implementation.
Table 3-6: Required RS-232 Options
Parameter Setting
Baud 2400
Byte 8 bits
Parity none
Stop Bits 1
Message Interval 2 seconds. When CRC opens
again, zero and span are done,
and the sample is being analyzed.
Range ID Relays: There are three dedicated Range ID relay contacts.
Three ranges are assigned to relays in ascending order—Low range is
assigned to Range 1 ID, Medium range is assigned to Range 2 ID, and
High range is assigned to Range 3 ID.
Network I/O: A serial digital input/output for local network protocol.
At this printing, this port is not yet functional. It is to be used in future
options to the instrument.
RS-232 Port: The digital signal output is a standard RS-232 serial
communications port used to connect the analyzer to a computer,
terminal, or other digital device. It requires a standard 9-pin D
connector.