Teledyne 3000TA Respiratory Product User Manual


 
Quickstart Guide Trace Oxygen Analyzer
Teledyne Analytical Instruments 16
Span Calibration
Make sure an appropriate span gas is connected to the sample port.
Auto Mode Spanning:
Press Span to enter the span function.
Use the arrow keys to toggle between AUTO and MAN span settling. Stop when AUTO
appears, blinking on the display.
Span: Settling: AUTO
<ENT> For Next
Press Enter.
Calib. Holding time
Cal hold: 5 min
Press Enter to move to the next screen.
Span Val: 000008.00
<ENT>Span <UP>Mod #
Use the ▲▼ arrow keys to enter the oxygen-concentration mode. The ◄► arrows
choose the digit, and the ▲▼ arrows choose the value of the digit. When you have
finished typing in the concentration of the span gas you are using, press Enter to
begin the span calibration.
#### ppm Span
Slope=#### ppm/s
The instrument automatically returns to the Analyze mode at the conclusion of the auto span
cycle..
Manual Mode Spanning:
Press Span to start the Span function.
Span: Settling:MAN
<ENT> For Next
Use the ▲▼ keys to toggle between AUTO and MAN span settling. Stop when MAN
appears, blinking, on the display. Press Enter to move to the next screen.