13
Functional Check
A functional check may be performed using the PROBE COM-
PENSATION c onnections on the front panel of the oscilloscope. See
Figure 3.
Figure 3: Probe functional check connections
1. Connect the probe to a TekConnect receptacle on the oscillo-
scope.
2. Set the oscilloscope to display the appropriate probe channel.
3. Using a ground lead and SMT KlipChip, connect the probe
ground to the outer ri ng of the PROBE COMPENSATION
connector on the oscilloscope.
4. Using a standard tip, hold the probe to the center conductor of the
PROBE COMPENSATION BNC connector on the oscilloscope.
5. Press AUTOSET, or adjust the oscilloscope to display a stable
calibration waveform. (You may need to adjust the probe offset to
display the waveform.)