Output (Fat Monitor Mode) Format
a Transmission Data
a Control Commands
Remote Mode
All types of data may be entered by connecting the BF-350 to a PC and switching to Remote mode.
When in Remote mode, the body fat analyzer may be operated remotely by sending control commands from a host.
a Error Codes
If an error occurs, the body composition analyzer will send an error code to the host.
a Switching to Remote Mode
The receiving PC must be ready to accept output data immediately after
measurement is completed.
ƽ
CAUTION
Output data
Byte Length
First byte 1
1
5
Second byte
K: unit of weight kg
L: unit of weight lb
S: stability of load (weight output)
D: instability of load (weight output)
I : overload status
--
-
: negative weight output
Header
Weight xxx.x
Output data format
¥ Terminal data are CR (ASCII format: 0DH), LF (ASCII format: 0AH).
¥ Measurement data will be sent in the format shown to the right:
x xxxx.x
1. When the unit is switched off, press the [ ]
and [ ] keys simultaneously.
ƽ
Note
The default setting is off.
2. Press [0], [8].
3. Switch Remote mode on and off by pressing [1]
or [0].
[1] :Enables Remote mode
[0] :Disables Remote mode
4. Press [ ] to confirm.
Only "r S" will appear (standby screen).
ƽNote
To switch off Remote mode, start
from the beginning again and
select "OFF" at Step 3.
ATHLETIC
STEP ON
FEMALE
MALE
kg
cm / ft.in
STABILIZED
AGE
% FAT
P T
ATHLETIC
STEP ON
FEMALE
MALE
kg
cm / ft.in
STABILIZED
AGE
% FAT
P T
ATHLETIC
STEP ON
FEMALE
MALE
kg
cm / ft.in
STABILIZED
AGE
% FAT
P T
ATHLETIC
STEP ON
FEMALE
MALE
kg
cm / ft.in
STABILIZED
AGE
% FAT
P T
*All fixed-length data right aligned. Enter for blanks.
0
U? :
Confirm current setting
U0: Use cm/kg mode
U1: Use in/lb mode
LFCR1U
LFCR0U
LFCR?U
commands
U
R? :
Confirm current setting
R0: Select equation for Westerners
R1: Select equation for Orientals
LFCR1R
LFCR0R
LFCR?R
commands
R
D? : Confirm current setting
D0 : Set clothing weight (5-byte fixed length)
D1 : Set gender male = 1 female = 2
D2 : Set body type standard = 0 athlete = 2
D3 : Set height (5-byte fixed length)
D4 : Set age (2-byte fixed length)
D6 : Set Target BF (2-byte fixed length)
LFCRXX6D
LFCRXX4D
LFCRXXXXX3D
LFCRX2D
LFCRX1D
LFCRX.XXX0D
LFCR?D
commands
D
G1: Start measuring body fat
G2: Measure weight only
LFCR2G
LFCR1G
commands
G
CHR$(01FH): Reset settings
LFCRCHR$(01FH)
Special
commands
E00 : Attempted to start measuring without completing settings
E01 : Error in calculating body fat percentage
EXX :other error ’ body fat analyzer switches off
LFCRXXE
LFCR10E
LFCR00E
*See p.26 "7. Troubleshooting" for details of errors.