ResMed C-Series Sleep Apnea Machine User Manual


 
11
English
Guidance and Manufacturer’s Declaration - Electromagnetic
Emissions and Immunity
Guidance and manufacturer’s declaration—electromagnetic emissions
The C-Series Tango system is intended for use in the electromagnetic environment specified
below. The customer or the user of the C-Series Tango system should assure that the device is
used in such an environment.
Emissions test Compliance
Electromagnetic
environment—guidance
RF emissions CISPR11 Group 1 The C-Series Tango system uses RF
energy only for its internal function.
Therefore, its RF emissions are very
low and are not likely to cause any
interference in nearby electronic
equ pment.
RF emissions CISPR 11 Class B The C-Series Tango system is suitable
for use in all establishments, including
domestic establishments and those
directly connected to the public low-
voltage network that supplies
buildings used for domestic purposes.
Harmonic Emissions
IEC 61000-3-2
Class A
Voltage Fluctuations/Flicker
Emissions IEC 61000-3-3
Complies
Medical Electrical Equipment needs special precautions regarding EMC and needs to be installed
and put into service according to EMC information provided in this document.
Warnings: The C-Series Tango system should not be used adjacent to or stacked with other
equipment. If adjacent or stacked use is necessary, the C-Series Tango system should be observed
to verify normal operation in the configuration in which it will be used.
The use of accessories (eg humidifiers) other than those specified in this manual is not
recommended. They may result in increased emissions or decreased immunity of the C-Series
Tango system.