Blender
Air-Oxygen
WARNING
An Oxygen Analyzer/Monitor must be used to verify
oxygen concentration.
Accuracy of oxygen concentration will be affected if bleed
is not activated at flow settings below 15 lpm for the High
Flow Blender, and 3 lpm for the Low Flow Blender.
DO NOT
obstruct the alarm.
DO NO
T use Blender when alarm is sounding.
DO NOT
use oil in or around the Blender.
DO NOT
occlude or obstruct the bleed port on the
auxiliary outlet of the Blender.
DO NOT
use near any type of flame or flammable/
explosive substances, vapors or atmosphere.
Oxygen Concentration Dial does not rotate 360 degrees.
Rotating the dial less than 21% or over 100% oxygen will
damage the Blender.
CAUTION
Turn off gas supplies when Air-Oxygen Blender is not in use.
Store the Air-Oxygen Blender in a clean, dry area when
not in use.
The Air-Oxygen Blender contains magnetic, ferrous
material that may affect the results of an MRI.
Ensure all connections are tight and leak free.
Avoid excessive pressure surges greater than 100 psi
(6.9 bar) when pressuring the Blender inlets.
DO NOT
steam autoclave.
DO NOT
immerse Air-Oxygen Blender into any liquid.
DO NOT
gas sterilize with (ETO) Ethylene Trioxide.
DO NOT
use if dirt or contaminants are present on or
around the Blender or connecting devices.
DO NOT
smoke in an area where oxygen is being
administered.
DO NOT
clean with aromatic hydrocarbons.