Precision Trading PM5300 Respiratory Product User Manual


 
11
Blender
Air-Oxygen
3. To activate the bleed, turn the knurled collar until the bleed
pin is engaged. Then slide the collar back until it contacts the
cover.
4. Confirm the concentration of air/oxygen with an Oxygen
Analyzer/Monitor.
5. Confirm the flow of air and/or oxygen mixture to the patient.
6. Turn “OFF” the Air and Oxygen supply when deactivating the
Auxiliary bleed, and no device is attached to the outlet port.
7. To deactivate the bleed, pull the collar out away from the cover
while rotating the collar until the slot engages the pin. Then
slide the collar past the pin and turn an additional
1
/4 turn to
lock in place.
NOTE: This Procedure is completed with ease when all gas
supplies to the Blender are turned “OFF”.
8. Turn “OFF” the Air and Oxygen supply or disconnect when
the Blender is not in use.
9. Operational Verification listed below should be done before
the Blender is placed in service.
Verification consists of:
• Alarm Test
Verify FiO
2
Concentration with
a Oxygen Analyzer/Monitor
• Reverse Gas Flow Procedure
PERFORMANCE CHECK
A detailed description of the tests listed below can be found in the
Blender Service Manual (P/N 504827), available on the Internet;
www.precisionmedical.com
• Reverse Gas Flow Procedure
• Operational Verification Procedure (OVP)
CLEANING
CAUTION
DO NOT
steam autoclave.
DO NOT
immerse the Air-Oxygen Blender into any liquid.
DO NOT
use any strong solvent or abrasive cleaners.
DO NOT
gas sterilize with (ETO) Ethylene Trioxide.
DO NOT
clean with aromatic hydrocarbons.