Oregon BPW128 Blood Pressure Monitor User Manual


 
12
GB
AUTOMATIC POWER OFF
The device will automatically switch off after about 1 minute.
However, to conserve battery power, press the button when
finished to turn the unit off.
ERROR MESSAGES
Display
Err 1
Err 2, 8,
12, 13,
14, 15
Err 3
Err 4
Err 5,
6, 7
Err 9,
10,11
Description
Deflation period too long
Incorrect measurement
Cuff is not fastened
Vigorous motion is being
detected during
measurement
Measurement error
Calibration error
Remedies
Do not move during deflation.
Check whether the unit is worn
well on wrist and measure
again.
Check if battery power is low.
Change all batteries and
measure again.
Fasten the cuff and measure
again.
Stay calm and easy then
measure again.
Stay calm and easy then
measure again.
Retry again, return the unit to
the manufacturer for repairing
if same symptom persists.
(Repairing/Replacement
charge free only if under the
valid warranty. Proof of
purchase is needed).