Omron Healthcare NE-C25 Nebulizer User Manual


 
The device is
abnormally loud.
Do not block the ventilation
slots located on the sides of the
compressor. Do not cover the
compressor with any type of
cover during use. Place the device
where the ventilation slots are not
obstructed during use.
The air filter cover is
incorrectly attached.
The device is very hot.
The ventilation slots are
blocked.
TROUBLESHOOTING GUIDE
22
PROBLEM SOLUTION
No power on unit when the
power switch is turned on.
Turn the power switch off.
Plug the power plug into an
electrical outlet. Turn the
device on.
No nebulization or low
nebulization rate when
the power is on.
Add the correct amount of
medication to the medication
cup.
Make sure the baffle is correctly
positioned over the nozzle in the
medication cup.
Make sure the nebulizer kit is
correctly assembled and the
inhalation accessory is correctly
attached.
Clean and disinfect the nebulizer
kit to remove the blockage.
Hold the nebulizer kit correctly.
Do not tilt the nebulizer kit so
the angle of the kit is greater
than 45 degrees.
Make sure the air tubing is
correctly attached to the
compressor and the nebulizer kit.
CAUSE
The device is not plugged
into an electrical outlet.
No medication in the
medication cup. Too much
or too little medication in
the medication cup.
The baffle is not placed
in the medication cup or is
incorrectly positioned.
The nebulizer kit is not
correctly assembled.
The nozzle is blocked.
The nebulizer kit is tilted
at an incorrect angle.
The air tubing is
incorrectly attached.
The air tubing is folded or
damaged. The air tube is
blocked.
Make sure the air tubing is not
folded, kinked or bent. Inspect the
air tubing for any damage. Replace
the air tubing if damaged.
The air filter is dirty.
Replace the air filter with a new
clean air filter.
Attach the air filter cover correctly.
Make sure the air filter cover is
not blocked.