PREPARING THE NEBULIZER FOR USE
5. Add the correct amount of prescribed medication to the
medication cup.
WARNING
For type, dose, and regime of medication follow
the instructions of your physician or licensed
healthcare practitioner.
WARNING
Do not spill water or other liquids on the compressor and power cord.
The compressor and power cord are not waterproof. If liquid spills on
these parts, immediately unplug the power cord and wipe off the liquid
with gauze or other soft absorbent material.
WARNING
Do not use tap or mineral water in the nebulizer kit for nebulizing purposes.
CAUTION
Do not add more than 6ml of medication to the medication cup.
6. Assemble the nebulizer kit by gently twisting the top and bottom
sections together clockwise. Make sure the two sections fit securely.