9
NOTE: The mesh cap may be damaged by hardened medication that is left in
the vibration horn effecting how the unit nebulizes. The following are maintenance
recommendations after each use to keep your unit working at its best level.
UNIT DISASSEMBLY & MAINTENANCE
NOTE: Be sure that the unit is turned OFF.
Remove the medication cup and dispose of any
remaining medications.
Be sure to nebulize all medication from the unit.
The medication cup needs to be removed while
the unit is on to remove all medication that
might be in the mesh cap or transducer. The unit
will no longer provide any aerosol when these
parts are completely empty.
NOTE: The lack of any misting will occur
with a high pitched sound when all medication
has been nebulized. Quickly turn the unit off
when this sound is heard it shows that there is
no more medication in the unit.
Pour a mixture of one part vinegar and three
parts distilled water into the medication cup
after each use. Nebulize the mixture for about
45 to 60 seconds. Remove the medication cup
and run the unit until it can no longer mist.
Turn the unit off at this point.
NOTE: Make sure that there is no medication
left on the mesh cap or horn that can dry and
cause damage or breakage to the mesh cap
when removed from the unit.
Place 2-3 drops of the water/vinegar mixture
onto the mesh cap surface when dried medica-
tion is noted. Nebulize distilled water from the
medication cup for at least 30 seconds.
We recommend that a small bottle of the
water/vinegar mixture be kept with your unit
at all times for proper cleaning.
1.
2.
3.