Omron ne-c801 Respiratory Product User Manual


 
9
EN
2. Guide to using the device
Preparing the Nebuliser for Use - Section 3 (page 10-13)
Warning:
Clean and disinfect the nebuliser kit, mouthpiece and nosepiece or masks
before using them for the first time after purchase, or if the device has not
been used for a long period of time, or if more than one person uses the
same device.
Taking Medication - Section 4
(page 14-16)
Caution:
• Do not tilt the nebuliser kit so that it is at an
angle of greater than 45
°.
• Do not crease or fold the air tube.
Cleaning and Daily Disinfecting
- Section 5 (page 17-21)
Note: Failure to clean the nebulising parts
after use may lead to infection.
How the nebuliser kit works
The medication that is pumped up
through the medication channel is
mixed with compressed air which is
generated by a compressor pump.
The compressed air mixed with
medication is turned into fine
particles and sprayed by being
impacted to the baffle.
Baffle
Aerosol
Medication
Channel
Medication
Compressed Air
Aerosol
Medication
Nozzle
Operating instructions