7
MX2 Basic English
Indication Error Cause Rectification
Correct Cuff pressure
reading low
not
possible Movement
Very weak
artery pulse
Cuff pressure
too high
Low Batteries weak/ Fit four new identical 1.5V
voltage exhausted batteries type AA
Display empty
• Use a soft, slightly moistened cloth to
clean the OMRON MX2 Basic.
Caution! Do not use petrol, thinners or
similar solvents!
• Use a soft, moistened cloth and soap to
clean the cuff.
Caution! Do not wash the cuff!
Caution! Do not use petrol, thinners or
similar solvents!
• Do not carry out repairs of any kind
yourself. If a defect occurs consult your
OMRON retail outlet or distributor as
mentioned on the packaging.
Calibration
• It is recommended to have the OMRON
MX2 Basic inspected every two years to
ensure correct function and accuracy.
Consult your OMRON retail outlet or
OMRON distributor as mentioned on the
packaging.
Replace batteries
• Remove the battery out of the battery
5 Error: causes and rectifications
6 Maintenance and spare parts
Wait for at least three minutes to repeat
measurement.
Inflate the cuff 40 mmHg higher than
previous value.
Follow instructions, repeat
measurement.
Wait for at least three minutes to repeat
measurement.
Inflate the cuff 30 mmHg lower than
previous value.
Follow instructions, repeat
measurement.
MX2 Basic 01-12 EN.qxd 10/7/08 12:19 PM Page 7